Precautions before using the constant temperature and humidity test box:
1. Confirm that the power cord is properly installed and properly grounded according to the specifications
2. Confirm whether the dry ball over-temperature protection switch is set at the maximum temperature of the device plus 10~20°C. This is the scope of the security structure of the device.
3. Check whether the wet bulb over-temperature protection switch is set to 150 °C
Note: The gauze should only be taken when the temperature test is done. If the gauze is running at a high temperature above 85°C, the gauze should be replaced before the next run. Otherwise, the water may not be able to absorb water. When changing the gauze, wash your hands first (gauze packaging When it is sterilized)
4. Confirm
- Is the amount of water sufficient?
- Is the tank lid properly closed?
- Is it clean? (Clean once a month, change gauze every three months)
5. Confirmation
-- Is the wet ball special gauze placed in the correct position? (Is it placed in front of the humidity sensor?)
- Is it wet? (Water is provided in the water tank. If the water level is normal, the gauze cannot be wetted. Please change the new gauze immediately. Change the gauze every three months.)
- Is the water tank normal?
6. Confirmation
- Is the humidifier water level normal? (Check the water level 3 to 5 minutes after the humidifier has supplied water)
- Is the water clean during humidification? (Humidifier should be drained regularly to keep the inside clean)
7. Confirm that the drain pipe is properly installed?