Malfunction and treatment method of salt spray test machine

Malfunction and treatment method of salt spray test machine

Shanghai Qingji Instrument Instrument Technology Co., Ltd. has been engaged in the research and development of salt spray testing machines for many years. The performance and operation of salt spray testing machines are briefly introduced. In order to allow users to quickly solve the problems caused by the fog aging test, the salt fog test machine can continue to operate normally. The following describes some common faults and treatment methods.

1. Reasons for temperature failure of salt spray test machine:

1. Because the temperature setting of the laboratory temperature controller is too low, you can set the temperature controller at the desired temperature a few;

2. Because the laboratory safety protection switch is set too low, the safety protection switch can be set to the required temperature;

3. For heating system failure, electromagnetic (electrical) failure, controller failure, please consult our after-sales department for debugging and repair.

2. Saturation barrel temperature failure of the salt spray test machine:

1. When the temperature of the saturation barrel temperature controller is too low, the temperature controller can be set to the desired temperature;

2. When the safety protection switch of the saturation barrel is set too low, the safety protection switch can be set to the required temperature;

3. Heating system failure, electromagnetic (electrical) failure, controller failure, please consult our after-sales department for debugging and repair.

3. Reasons for insufficient spray volume of salt spray tester:

1. When the spray regulator is placed too low, the spray regulator can be adjusted up;

2. The glass filter in the preheating tank is blocked, and the glass filter can be cleaned;

3. When the pressure setting is too low, the pressure regulating valve can be adjusted to a pressure of 1KG per square centimeter. The air compressor is marked with a regulating valve to adjust to a pressure of 2KG per square centimeter.

4. When the low water level alarm light of the machine is on, it means that the water level is too low. It is necessary to check whether the water level of the test room or the saturation tank is sufficient.

5. When the air compressor is not running, the air compressor itself has the function of self-contained, and can be used as usual at this time, without affecting the test process of the test.

6. When the water tank is not running, the power supply will be cut off when the water level in the heating tank is too low, and the water level in the heating tank can be increased to normal conditions.

7. When the temperature display or the temperature sensor fails, please consult our after-sales department for debugging and repair.

Diy Diamond Painting

Diamond Ring has provided consumers with glue for diamond setting on the bottom map of diamond setting, which does not require consumers to use additional glue, which improves the safety and efficiency of operation, making it possible for 6-year-old children to 60-year-old sixty-year-old elderly. Can be set with diamonds. Many customers tell us that DIY Diamond Painting makes complicated things so simple.